0

Out of 0 Ratings

Owner's of the 3M Marine Instruments Bair Hugger Temperature Management Unit gave it a score of 0 out of 5. Here's how the scores stacked up:
  • Reliability

    0 out of 5
  • Durability

    0 out of 5
  • Maintenance

    0 out of 5
  • Performance

    0 out of 5
  • Ease of Use

    0 out of 5
of 19
 
10 English Bair Hugger Model 505 Temperature Management Unit Operator’s Manual
3M Patient Warming 3M Health Care.
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202431A
Mounting the Model 505 Temperature Management Unit
   
The Model 505 unit clamps easily to an IV pole (see Figure E). Simply turn the handle clockwise to
tighten the clamp onto an IV pole, counterclockwise to release.
Warning: To prevent tipping, clamp the Model 505 temperature management unit to an IV pole at
a height that provides stability. We recommend clamping the unit no higher than 44 in. (112 cm)
from the floor on an IV pole with a minimum 28 in. (71 cm) diameter wheelbase. Failure to do so
may result in IV pole tipping, catheter site trauma, and patient injury.
  
The Model 505 unit can also hang on the edge of a bed. The safety strap is designed to loop
around the bedrail, keeping the Model 505 unit safely suspended even if the unit is inadvertently
dislodged from the bedrail (see Figure F).
≥28 in. (71 cm)
≤44 in.
(112 cm)
Figure E. Model 505 unit attached to an IV pole Figure F. Model 505 unit attached to a bedrail
Safety strap